Comparison

Human Aminopeptidase N/ANPEP ELISA Kit

Item no. ELH-ANPEP-1
Manufacturer Raybiotech
Amount 1 Plate Kit
Quantity options 1 Plate Kit 2 Plate Kit 5 Plate Kit
Category
Type Elisa-Kit
Specific against Human
Host Hamster - Armenian
Sensitivity 40 pg/ml
ECLASS 10.1 32160605
ECLASS 11.0 32160605
UNSPSC 41116126
Available
Specificity This ELISA antibody pair recognizes Human Aminopeptidase N.
Compatible Sample Types
Cell Culture Supernatants
Plasma
Serum
Design Principle
Sandwich-based
Method of Detection
Colorimetric
Quantitative/Semi-Quantitative
Quantitative
Solid Support
96-well Microplate
Gene Symbols
ANPEPAPNCD13PEPN
Kit Components
Pre-
Coated 96-well Strip Microplate
Wash Buffer
Stop Solution
Assay Diluent(s)
Lyophilized Standard
Biotinylated Detection Antibody
Streptavidin-
Conjugated H
R
P
T
M
B One-
Step Substrate
Other Materials Required
Distilled or deionized water
Precision pipettes to deliver 2 µl to 1 µl volumes
Adjustable 1-25 µl pipettes for reagent preparation100 µl and 1 liter graduated cylinders
Tubes to prepare standard and sample dilutions
Absorbent paper
Microplate reader capable of measuring absorbance at 450nm
Log-log graph paper or computer and software for E
L
I
S
A data analysis
Protocol Outline
Prepare all reagents, samples and standards as instructed in the manual.
Add 100 µl of standard or sample to each well.
Incubate 2.5 h at R
T or O/
N at 4°
C.
Add 100 µl of prepared biotin antibody to each well.
Incubate 1 h at R
T.
Add 100 µl of prepared Streptavidin solution to each well.
Incubate 45 min at R
T.
Add 100 µl of T
M
B One-
Step Substrate Reagent to each well.
Incubate 30 min at R
T.
Add 50 µl of Stop Solution to each well.
Read at 450 nm immediately.
Need your results faster? Try Ray
Biotech's Speed
E
L
I
S
A platform for quantitative detection in just three hours.
Protein Name & Synonyms
Aminopeptidase N (AP-N) (hAPN) (EC 3.4.11.2) (Alanyl aminopeptidase) (Aminopeptidase M) (AP-M) (Microsomal aminopeptidase) (Myeloid plasma membrane glycoprotein CD13) (gp150) (CD antigen CD13)
Recommended Dilution (serum & plasma)
1, 000 fold
Additional Notes
Strip plates and additional reagents allow for use in multiple experimentsQuantitative protein detectionEstablishes normal rangeThe best products for confirmation of antibody array data

Note: The presented information and documents (Manual, Product Datasheet, Safety Datasheet and Certificate of Analysis) correspond to our latest update and should serve for orientational purpose only. We do not guarantee the topicality. We would kindly ask you to make a request for specific requirements, if necessary.

All products are intended for research use only (RUO). Not for human, veterinary or therapeutic use.

Amount: 1 Plate Kit
Available: In stock
available

Compare

Add to wishlist

Get an offer

Request delivery time

Ask a technical question

Submit a bulk request

Questions about this Product?