Comparison

CLIA Kit for Angiotensinogen (AGT)

Manufacturer Cloud-Clone
Category
Type Clia
Specific against Human
Amount 24T
Item no. SCA797Hu-24T
eClass 6.1 27150207
eClass 9.0 27150207
Available
Alias ANHU; SERPINA8; Angiotensin; Renin Substrate; Serpin Peptidase Inhibitor,Clade A,Member 8(Alpha-1 Antiproteinase,Antitrypsin)
Quantity options
Detection range
0.14-100ng/mL
Organism species
Homo sapiens (Human)
Sensitivity
The minimum detectable dose of this kit is typically less than 0.06ng/mL
Sample type
Serum, plasma and other biological fluids
Assay length
2h, 40min
Method
Double-antibody Sandwich
Specificity

This assay has high sensitivity and excellent specificity for detection of Angiotensinogen (AGT).
No significant cross-reactivity or interference between Angiotensinogen (AGT) and analogues was observed.

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Angiotensinogen (AGT) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Angiotensinogen (AGT) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary
1. Prepare all reagents, samples and standards;
2. Add 100µ L standard or sample to each well. Incubate 2 hours at 37° C;
3. Aspirate and add 100µ L prepared Detection Reagent A. Incubate 1 hour at 37° C;
4. Aspirate and wash 3 times;
5. Add 100µ L prepared Detection Reagent B. Incubate 30 minutes at 37° C;
6. Aspirate and wash 5 times;
7. Add 100µ L Substrate Solution. Incubate 10 minutes at 37° C;
8. Read RLU value immediately.
Test principle
The microplate provided in this kit has been pre-coated with an antibody specific to Angiotensinogen (AGT). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Angiotensinogen (AGT). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Angiotensinogen (AGT) level in the sample or standard.;
Research Area
Cardiovascular biology;

Note: The presented information and documents (Manual, Product Datasheet, Safety Datasheet and Certificate of Analysis) correspond to our latest update and should serve for orientational purpose only. We do not guarantee the topicality. We would kindly ask you to make a request for specific requirements, if necessary.

All products are intended for research use only (RUO). Not for human, veterinary or therapeutic use.

Amount: 24T
Available: In stock
available

Compare

Add to wishlist

Get an offer

Request delivery time

Ask a technical question

Submit a bulk request

Questions about this Product?
 
Close