Vergleich

CLIA Kit for Hemoglobin (HB)

ArtNr CCB409Hu-24T
Hersteller Cloud-Clone
Menge 24 T
Quantity options 10x96 T 24 T 48 T 5x96 T 96 T
Kategorie
Typ Clia
Applikationen IA
Specific against Human (Homo sapiens)
Sensitivity The minimum detectable dose of this kit is typically less than 0.42ug/mL
ECLASS 10.1 32161000
ECLASS 11.0 32161000
UNSPSC 41116126
Alias Hgb; Haemoglobin; Heterotetramer(αβ)2
Lieferbar
Specificity Homo sapiens (Human)
Quantity options
Detection range
1.17-300ug/mL
Organism species
Homo sapiens (Human)
Sensitivity
The minimum detectable dose of this kit is typically less than 0.42ug/mL
Sample type
Serum, plasma, tissue homogenates, erythrocyte lysates, cell culture supernates and other biological fluids
Assay length
2h
Method
Competitive Inhibition
Specificity

This assay has high sensitivity and excellent specificity for detection of Hemoglobin (HB).
No significant cross-reactivity or interference between Hemoglobin (HB) and analogues was observed.

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Hemoglobin (HB) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Hemoglobin (HB) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary
1. Prepare all reagents, samples and standards;
2. Add 50µ L standard or sample to each well.
        And then add 50µ L prepared Detection Reagent A immediately.
        Shake and mix. Incubate 1 hour at 37° C;
3. Aspirate and wash 3 times;
4. Add 100µ L prepared Detection Reagent B. Incubate 30 minutes at 37° C;
5. Aspirate and wash 5 times;
6. Add 100µ L Substrate Solution. Incubate 10 minutes at 37° C;
7. Read RLU value immediately.
Test principle
The microplate provided in this kit has been pre-coated with a monoclonal antibody specific to Hemoglobin (HB). A competitive inhibition reaction is launched between biotin labeled Hemoglobin (HB) and unlabeled Hemoglobin (HB) (Standards or samples) with the pre-coated antibody specific to Hemoglobin (HB). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Hemoglobin (HB) in the sample. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is reverse proportional to the Hemoglobin (HB) level in the sample or standard.
Research Area
Metabolic pathway; Infection immunity; Hematology; Neuro science;

Hinweis: Die dargestellten Informationen und Dokumente (Bedienungsanleitung, Produktdatenblatt, Sicherheitsdatenblatt und Analysezertifikat) entsprechen unserem letzten Update und sollten lediglich der Orientierung dienen. Wir übernehmen keine Garantie für die Aktualität. Für spezifische Anforderungen bitten wir Sie, uns eine Anfrage zu stellen.

Alle Produkte sind nur für Forschungszwecke bestimmt. Nicht für den menschlichen, tierärztlichen oder therapeutischen Gebrauch.

Menge: 24 T
Lieferbar: In stock
lieferbar

Vergleichen

Auf den Wunschzettel

Angebot anfordern

Lieferzeit anfragen

Technische Frage stellen

Bulk-Anfrage stellen

Fragen zum Produkt?
 
Schließen