ArtNr |
BM-EKU11555 |
Hersteller |
Biomatik
|
Menge |
96T |
Kategorie |
|
Typ |
Elisa-Kit |
Applikationen |
CLIA |
Specific against |
Mouse |
ECLASS 10.1 |
32160605 |
ECLASS 11.0 |
32160605 |
UNSPSC |
41116126 |
Lieferbar |
|
Format |
96-Well Plate |
Target Name |
Anti-Mullerian Hormone |
Target Synonyms |
MIF; MIH; MIS; Müllerian Inhibiting Factor; Müllerian Inhibiting Hormone; Müllerian Inhibiting Substance |
Sample type |
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
Detection ra nge |
39.06-10, 000 pg/mL |
Sensitivity |
14.13 pg/mL |
Precision |
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Anti-Mullerian Hormone (AMH) were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Anti-Mullerian Hormone (AMH) were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/meanX100. Intra-assay: CV<10%. Inter-Assay: CV<12%. |
Detection Method |
Chemiluminescent |
Assay Time |
2 hours |
Assay Type |
Competitive Inhibition |
Shipping Condition |
Ice packs |
Storage |
Short term: 4CC; Long term: see manual. |
Precaution of Use |
The Stop Solution is acidic. Do not allow to contact skin or eyes. |
Expiry Date |
8 months |
UniProt ID |
n/a |
Specificity |
This assay has high sensitivity and excellent specificity for detection of Anti-Mullerian Hormone (AMH). |
Cross Activity |
This assay has high sensitivity and excellent specificity for detection of Anti-Mullerian Hormone (AMH). |
Test principle |
The microplate provided in this kit has been pre-coated with a monoclonal antibody specific to Anti-Mullerian Hormone (AMH). A competitive inhibition reaction is launched between biotin labeled Anti-Mullerian Hormone (AMH) and unlabeled Anti-Mullerian Hormone (AMH) (Standards or samples) with the pre-coated antibody specific to Anti-Mullerian Hormone (AMH). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Anti-Mullerian Hormone (AMH) in the sample. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is reverse proportional to the Anti-Mullerian Hormone (AMH) level in the sample or standard. |
Assay procedure summary |
1. Prepare all reagents, samples and standards; 2. Add 50 uL standard or sample to each well. And then add 50 uL prepared Detection Reagent A immediately. Shake and mix. Incubate 1 hour at 37CC; 3. Aspirate and wash 3 times; 4. Add 100 uL prepared Detection Reagent B. Incubate 30 minutes at 37CC; 5. Aspirate and wash 5 times; 6. Add 100 uL Substrate Solution. Incubate 10 minutes at 37CC; 7. Read RLU value immediately. |
Stability |
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end. |
Restrictions |
For Research Use Only. Not for use in diagnostic procedures, drug use, or for administration to humans or animals. |
Quality Systems |
The kit is manufactured at ISO 9001 and ISO 13485 certified facilities. |
Quality Assurance |
Biomatik ELISA Kits are not typically pre-assembled as finished products due to limited shelf life. Before shipping, each kit is assembled and tested to ensure that it meets specifications before shipping. Minor changes may occur to the Detection Range, Sensitivity, and Precision. All kits are tested to confirm that they fall within their defined Inter- and Intra- assay coefficient of variation. Please always refer to the hard copy manual included in The kitbefore experiment. |
Hinweis: Die dargestellten Informationen und Dokumente (Bedienungsanleitung, Produktdatenblatt, Sicherheitsdatenblatt und Analysezertifikat) entsprechen unserem letzten Update und sollten lediglich der Orientierung dienen. Wir übernehmen keine Garantie für die Aktualität. Für spezifische Anforderungen bitten wir Sie, uns eine Anfrage zu stellen.
Alle Produkte sind nur für Forschungszwecke bestimmt. Nicht für den menschlichen, tierärztlichen oder therapeutischen Gebrauch.